112
Detailed information about your system
9
Electrical and safety standards
This meter complies with CISPR 11: Class B (Radiated 
Only). Emissions of the energy used are low and not likely 
to cause interference in nearby electronic equipment. 
The meter has been tested for immunity to electrostatic 
discharge as specified in IEC 61326-2-6. This meter 
complies with immunity to radio frequency interference as 
specified in IEC 61326-1 and 61326-2-6.
The meter meets the requirements for immunity to 
electrical interference at the frequency range and test level 
specified in international standard ISO 15197.
Use of this meter near electrical or electronic equipment 
that are sources of electromagnetic radiation, may 
interfere with proper operation of this meter. It is 
advisable to avoid testing in close proximity to sources of 
electromagnetic radiation.
Do Not use the equipment where aerosol sprays are being 
used, or when oxygen is being administered.
in
PF3130451Rev1_OTSPF_OB_I_GB_en_zug_R2.indd   112 3/12/15   11:36 AM
 Statement of Use: Verify status before each use
Effectivity Date: Mar 18 2015       Status: Current     State: Released
Type: Project File  Name: 3130451  Revision: 1