106
Detailed information about your system
9
System Accuracy Results for Glucose 
Concentrations *5.55mmol/L
Percent (and number) of meter results that match the 
laboratory test
Within ±5%  Within ±10%  Within ±15% 
63.3%
(266/420)
95.5%
(401/420)
99.8%
(419/420)
System Accuracy Results for Glucose 
Concentrations between 1.5mmol/L 
and 26.8mmol/L
Percent (and number) of meter results that match the 
laboratory test
Within ±0.83mmol/L or ±15%
98.8% (593/600)
NOTE: Where 1.5mmol/L represents the lowest glucose 
reference value and 26.8mmol/L represents the highest 
glucose reference value (YSI value).