System accuracy results for glucose
concentrations ≥5.55mmol/L
Percent (and number) of meter results that match the
laboratory test
Within ±5% Within ±10% Within ±15%
65.2% (270/414) 93.2% (386/414) 99.3% (411/414)
System accuracy results for glucose
concentrations between 1.9mmol/L and
25.0mmol/L
Within ±0.83mmol/L or ±15%
99.5% (597/600)
NOTE: Where 1.9mmol/L represents the lowest glucose
reference value and 25.0mmol/L represents the highest
glucose reference value (YSI value).