160
12
System information
Regression statistics
Samples were tested in duplicate on each of three test
strip lots. Results indicate that the OneTouch Verio
Reflect
System compares well with a laboratory method.
# of
Subjects
# of Tests Slope Intercept
(mmol/L)
100 600 1.02 0.04
95% CI
Slope
95% CI
Intercept
(mmol/L)
Std. Error
(S
y.x
)
(mmol/L)
R
2
1.01 to 1.03 -0.03 to 0.11 0.49 0.99