162
12
System information
Precision
Within Run Precision (300 Venous Blood Samples Tested
per Glucose Level)
Data generated using the OneTouch Verio Reflect
Meter.
Target
Glucose
(mmol/L)
Mean
Glucose
(mmol/L)
Standard
Deviation
(mmol/L)
Coecient
of Variation
(%)
2.2 2.2 0.06 2.68
5.0 4.9 0.13 2.66
7.2 7.1 0.21 2.98
11.1 11.2 0.31 2.73
19.4 19.8 0.52 2.60
Results show that the greatest variability observed
between test strips when tested with blood is
0.13mmol/L SD or less at glucose levels less than
5.55mmol/L, or 2.98% CV or less at glucose levels at
5.55mmol/L or above.