F1 F2 F3 F4
Press FLO (F2):
F1 F2 F3 F4
Press UNT (F2) for Unit selection.
The screen will show:
F1 F2 F3 F4
Press NXT (F1) to change selection and OK (F4) to accept.
NOTE: The totalizer (total flow measured) will roll over when reaching a certain value.
The maximum value is dependent on the flow units selected (see Totalizer Rollover p.
56).
Selections for flow units are:
SCFM LBS/M MSCFD (MCFD)
SCFH LBS/S SM3/H
NM3/H NLPH MT/H
NM3/M NLPM NM3/D
KG/H MMSCFD (MMCFD) MMSCFM
KG/M LBS/D SCFD
KG/S SLPM MCFD (MSCFD)
LBS/H NLPS SM3/M
SM3/D
WARNING!
The F-5500 re-calculates area, 4 and 20mA values, maximum flow for the pulse output
and zero flow cutoff when changing flow units.
NOTE: The totalizer values do not adjust or correct for the new units. When changing
flow units, log the totals in the totalizer and then reset the totalizer after they’ve changed
flow units to maintain accuracy in the totalizer.
11451 Belcher Road South, Largo, FL 33773 • USA • Tel +1 (727) 447-6140 • Fax +1 (727) 442-5699 • sales@onicon.com
F-5500 Thermal Mass Flow Meter Manual 06/17 - 2030 / 107023 Page 49
Model F-5500
OPERATION
Operation: Programming
Programming
by Display
MAIN MENU
I/O FLO DSP EXIT
FLOW MENU 1
DGN UNT FM2 EXIT
FLO = SCFH
NXT OK
Flow Units