F-5500 Thermal Mass Flow Meter Manual 06/17 - 2030 / 107023 Page 67
Model F-5500
COMMUNICATIONS
Model F-5500
COMMUNICATIONS
Modbus Programming
Communication Protocol and Parameters
To program the communication parameters, start at the Main Menu:
F1 F2 F3 F4
Then press I/O (F1) to set Inputs/Outputs:
F1 F2 F3 F4
Then press COM (F1) to select communication parameters.
Set Bus protocol for Modbus:
F1 F2 F3 F4
Press NXT (F1) repeatedly until Modbus is selected as shown and then press OK (F4)
to accept the setting.
The following communication parameters are only available for MODBUS:
F1 F2 F3 F4
Press NXT (F1) repeatedly until the correct selection is shown and then press OK (F4)
to accept the setting.
Baud=9600
NXT OK
SET I/O
COM 420 EXIT
Comm=MODBUS
NXT OK
Modbus
Communication
Parameters
MAIN MENU
I/O FLO DSP EXIT
This is the Main Menu for the programming mode. To exit the programming mode,
press EXIT (F4) repeatedly until “Normal Mode” is seen briefly.