Maintenance
- 32 -
optek-Manual--1004-2003-02--AF45-US-2011-01-04
www.optek.com
9.4 Return to operation after the exchange of lamp module and
detector module
Before return to operation of the entire measuring system, undertake the follow-
ing activities:
1. Switch on the converter.
2. Wait for approx. 15 minutes, until the converter has reached the working
temperature.
3. Check the system's zero-point (see instruction manual of the converter).
4. Document your settings using the form (see instruction manual of the con-
verter).
5. Check measuring results with regard to plausibility.
6. If settings and measuring results are correct, enable measuring.