Installation
- 19 -
optek-sensors AS56, Version 04.2010_1.2US, 15.04.2010
www.optek.com
Install sensor in vertical pipelines!
Orientate port to one side in horizontal pipelines!
Check seat of swivel nut during installation and maintenance activities.
Do not exceed 25 bar during any pressure test.
Always install weld-in port, adapter or T-piece in the pipeline providing
optimum flow through the measuring segment. Do not install the named
parts immediately behind valves or other flow obstacles.
Make sure that no extraneous light sources are located near the sensor
in order not to distort measuring results.
Only remove a sealing flange, if the pipeline is pressureless, cooled down
and empty. The sealing flange is not suitable for safety ports.
In rare cases seizing of threads may occur. An appropriate lubricant has
thus to be used. Apply lubricant as a thin film to the male thread of the
port. Appropriate lubricants are given in the spare parts list.