8.3 PHACOEMULSIFICATION
User Interface
Main U/S box
Reduction Mode U/S box
Parameter windows available
- FLOW (aspiration flow)
- VAC (limit vacuum)
- POWER (U/S energy)
Irrigation parameters
- Status (ON/OFF)
- I.V. Pole height
Aspiration parameters
- Vacuum mmHg (5 – 500)
- Flow rate cc/min (2-50)
- RiseTime cc/min (2-50)
- Mode (linear/panel)
Ultrasound parameters
- Phaco tip stroke µm (5 –
100)
- Mode (linear/panel)
- U/S Rate (Continuous,
Short Pulse, Single Burst,
Multi-Burst, Continuous
Burst)
- Pulse frequency
(1,2,3,4,5,6,8,10,13,16,20,
25,32,40 bursts/sec)
- Timer (reading and reset
key)
- Equivalent Phaco Time
Touching the U/S Function Key will display U/S memory 1 screen (U/S 1). In order
to display U/S Memory 2 screen (U/S 2), touch the <B> key. Further depression
activates U/S 3. Swapping between memories U/S 1, U/S 2 and U/S 3 can also be
achieved by means of the footswitch. For footswitch operation please refer to the
relevant section.