PULSAR Minimal Stress Installation and operating Manual
OPTIKON 2000
Cod. 111001EN 2012-02-16 Rev.J
10-1
10. TROUBLESHOOTING GUIDE
10.1 WARNINGS
When the "U/S" Ultrasound function is selected, the following messages may appear
in the status box in the centre of the screen:
ENGLISH ITALIANO
MESSAGE MESSAGGIO
1 Plug handpiece Collegare il manipolo
2 Please prime… Inizializzare il sistema
3 Check tip Controllare la punta
4 Phaco ready Ultrasuoni pronti
5 Weak handpiece Manipolo poco efficiente
10.2 MISCELLANEOUS PROBLEMS
The troubleshooting guide lists some of the anomalies that may occur in the unit’s
operation, the related symptoms and corrective actions. If the problems with the
PULSAR surgical system persist even after having carried out the corrective actions
below, contact an OPTIKON 2000 authorised service centre.
SYMPTOMS CORRECTIVE ACTION
Mains power supply
PULSAR completely inoperative: a) A.C. power cord not connected. Connect power
cord to control console and the mains.
b) A.C. line fuse blown. Replace it.