PULSAR Minimal Stress Installation and operating Manual
OPTIKON 2000
Cod. 111001EN 2012-02-16 Rev.J
9-3
CAUTION
1) Do not use synthetic detergents or oil-based soaps
2) Be sure that the diathermy handpiece is completely dry before
using it, as RF currents may find a path through wet surfaces.
3) Vitrectomy handpieces should never be completely immersed in
distilled water as the water may enter the high pressure membrane
chamber thus causing malfunction.