EasyManua.ls Logo

Otto Bock ActiGait - 1. Introduction; 2. Intended Use; 2.1 Medical Purpose; 2.2 Application

Otto Bock ActiGait
69 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3
1. Introduction
This manual describes the use of the ActiGait® and is directed towards individuals
with an ActiGait® Implant. The manual is to be used as an information resource and
an instruction manual. The term “User” denes individuals who have an ActiGait®
Implant and thereby are users of the ActiGait®.
For clinicians the Clinician Manual (647G805=GB) is available
2. Intended Use
2.1 Medical Purpose
The ActiGait® is solely to be used as treatment in patients having a drop foot follow-
ing an upper motor neuron lesion.
2.2 Application
The ActiGait® system is a partially implantable medical device for treatment of drop
foot which activates the muscles of the lower leg during walking. The system is
intended for use by persons suering from paralysis of the ankle dorsiexor muscles
caused by damage to the central nervous system. The largest group of potential users
are persons with hemiplegia as a consequence of stroke. The system will not function
for people with drop foot caused by peripheral nerve damage.
2.3 QualicationofClinician
The selection of users for ActiGait®, setup and adjustment are done by clinicians such
as medical doctors, physiotherapists or equivalent. To be able to program and adjust
the ActiGait®, the clinician must have obtained a certicate on usage of the Clinical
Station software following training by Otto Bock
2.4 Follow up Schedule
It is recommended that the ActiGait® is activated 1-3 weeks after implantation, as the
wound healing must be nished rst. The user should thereafter be followed closely
with 2-3 follow-ups during the rst 6 months for ne-tuning of the settings and gen-
eral guidance. Subsequently, the user should be seen in the clinic on request.

Table of Contents