User’s Manual of OWON Colour Digital Storage Oscilloscope
85
Record length
z 5K points on each channel---- PDS5022S
z 6K points on each channel
---- PDS6042S, PDS6062S, PDS6062T,
PDS7062T, PDS7102T
Scanning speed
(S/div)
●5ns/div~100s/div, step by 1~2.5~5
---- PDS5022S
●5ns/div~100s/div, step by 1~2~5
---- PDS6042S, PDS6062S, PDS6062T,
PDS7062T, PDS7102T
Sampling rate / relay
time accuracy
±100ppm
Interval(△T)
accuracy(DC~
100MHz)
Single:±(1 interval time+100ppm×
reading+0.6ns)
Average>16:±(1 interval time +100ppm×
reading+0.4ns)
A/D converter 8 bits resolution (2 Channels simultaneously)
Sensitivity
5mV/div~5V/div(at input)
Displacement
●±10div(5mV/div~5V/div)----PDS5022S
●±50V(500mV~5V),±2V(5mV~200mV)
---- PDS6042S, PDS6062S, PDS6062T,
PDS7062T, PDS7102T
Single bandwidth Full bandwidth
Low Frequency
≥5Hz (at input, AC coupling, -3dB)
Rise time
(at input, Typical)
≤14ns----PDS5022S
≤8.75ns---- PDS60462S
≤5.8ns---- PDS6062S, PDS6062T, PDS7062T
≤3.5 ns----PDS7102T
Vertical
system:
DC accuracy
±3%