Baseline Drift, Continued
Monitor high-voltage for
stability. Replace high-voltage
board if necessary.
Use Light Leak Check
procedure to check. Correct if
necessary.
Baseline High-Frequency
Noise
Determine cause of instability
and correct.
High-voltage power supply is
unstable
Use high-voltage adjustment
procedure to monitor high-
voltage for stability. Replace if
necessary.
Baseline Low-Frequency
Noise
Use light leak test to find leaks
and repair as needed.
Determine cause of instability
and correct.
Use light leak test to find leaks
and repair as needed.
Use gas leak test to find leaks
and repair as needed.
Incomplete sample
combustion
Ensure furnace temperature
and gas flows are correct.