Unstable Sulfur Detector
Response
Unstable PMT high-voltage
Adjust or replace high-voltage
PCB.
Use gas leak test to find leaks
and repair as needed.
Use light leak test to find leaks
and repair as needed.
Adjust or replace high-voltage
supply or cable.
Use light leak test to find leaks
and repair as needed.
Defective High-Voltage PS
Contaminated pyrotube,
transfer tubes or fluorescence
chamber
Clean or replace pyrotube,
transfer tubes or fluorescence
chamber.
Check flow system for leaks
and repair as needed. See
““Gas Leak Testing”,
Chapter 4.
Contaminated Detector (Odor
and discoloration inside
detector)
Remove and clean detector.
Contaminated pyrotube or
fluorescence chamber
Clean or replace pyrotube, or
fluorescence chamber.
Water condensed in detector
or flow system
Sample zero or other dry gas
overnight. Proper response
will return after the water is
removed.
Use gas leak test to find leaks
and repair as needed.
Contaminated Flow System
(Drift stops after gas flows are
turned OFF)
Clean or replace tubing in flow
system as needed.