Page 74 of 92 PAC L.P., All Rights Reserved, 2015
R6000S Operation Manual: Rev. G
Reduce Sample size,
concentration or amplifier gain.
Clean or replace Pyrotube.
Incomplete sample
combustion
Ensure furnace temperature
and gas flows are correct.
Check fittings, ferrules,
Pyrotube, and vents in the
flow path and remove
restrictions.
Replace or adjust PMT HV.
Abnormal ambient temp
change
Stabilize ambient temperature.
Use light leak test to find leaks
and repair as needed.
Defective Pre-amp PCB or
Amplifier PCB
Replace Pre-amp PCB or
Amplifier PCB.
Use gas leak test to find leaks
and repair as needed.
Inaccurate calibration curve
Perform Calibration
Procedure.
Inaccurate calibration
standards
Clean or replace Pyrotube.
Incomplete sample
combustion
Ensure furnace temperature
and gas flows are correct.