Page 76 of 92 PAC L.P., All Rights Reserved, 2015
R6000S Operation Manual: Rev. G
Symptom
Probable Cause
Corrective Action
No Detection
No high-voltage
Replace high-voltage supply.
Soot in System
Check for soot deposits in
system.
Gas leak
Use gas leak test to find leaks
and repair as needed.
No detector electronics
response
Cause a light leak in the
detector to determine
operation of electronics.
Defective Supply
Check all supply outputs and
replace defective supply.
Membrane Dryer leaking
internally
Replace Membrane Dryer.
Water in Reaction Chamber
Remove all water from
chamber.
Moisture Collected In
Detector or Flow System
Tubing
Membrane dryer(s) is bad
Replace membrane dryer.
Dryer flow is too low
Check dryer flow meter and
increase flow as necessary.