K3.1.142k Manual NitroFlow Lab - 14 -
5 Installation
Follow the paragraphs in this chapter to install the generator.
5.1 Transport
Warning
x Transport the generator upright.
x Put the generator in the original box to transport the generator over longer distances.
x Lift the generator with a forklift.
x For qualifications of personnel, refer to §2.1.
5.2 Define location
IMPORTANT
x The generator contains compressors that generate heat; for optimal performance and
lifetime it is necessary that cooling air can be vented without resistance. A minimum
clearance distance from walls or other objects of at least 50 cm/ 20 inches on all sides
(back, left, right and top) is a necessity; also efficient local ventilation at the ventilation
outlet is highly recommended especially when the device is installed under a bench
Install the generator on a fixed location. The location must meet the following requirements:
x Minimum clearance of 50 cm on all sides (back, left, right and top) as to facilitate heat removal
x Indoors
x Dry
x No continuous direct irradiation by sunlight
x Away from heat sources
x Properly ventilated room.
x Easy accessibly for operating and service
5.3 Unpack and check equipment
x Open the packaging per instructions on the crate.
x Make sure that all components are delivered. Refer to § 4.1.
5.4 Connect nitrogen consumer
Warning
x Do not connect the power at this time.
x Make sure that the inlet and outlet tubes are free of dust, particles, metal parts and curls,
liquids and grease before you connect the generator.
Connect the product outlet to the application.
5.5 Connecting power
1. Connect the mains plug to a suitable wall socket with earth connection
2. The control system has input and output contacts for remote control and alarm signaling (refer to §5.6).
Warning
The main supply line voltage must be within 10% of nominal rated voltage for the generator. In
case of larger variations the generator will stop; continued use under these circumstances will
inevitably lead to motor damage.