K3.1.142k Manual NitroFlow Lab - 18 -
6.3.1 Log on menu
Access: Touch log on menu button in settings screen (refer to §6.3)
ATTENTION: When you start-up the system for a first time you do not need to enter a PIN
CODE
Function: Protect the settings in the system with a (personal) pin code.
In the log on menu:
x Enter the default pin code (1234) after selecting PINCODE YES under the options menu (refer to §6.3.4).
x Change the default pin code to a personal pin code of 4 digits (refer to §6.3.4)
x Return to default factory settings by entering pin code 7833 (refer to §4.1)
x In case you lost your pin code, please contact your supplier
Caution:
When returning to factory settings, the alarms, p-switch, options and settings must be reset.
Also the log on pincode is back to default value 1234
6.3.2 Alarm settings menu
Access: Touch alarm settings menu button in settings screen (see § 6.3.2)
Function: Set different alarms
In the alarm settings menu it is possible to set 6 different alarms.
Screen Alarm Explanation
1/6 O2 high oxygen level too high
2/6 O2 low oxygen level too low
3/6 Pres. Inlet high inlet pressure too high
4/6 Pres. Inlet high inlet pressure too high
5/6 Pres. Outlet high outlet pressure too high
6/6 Pres. Outlet low outlet pressure too low
1. To activate an alarm touch button A. When the button is touched you can select the options YES, AUTO
RESET or NO by pressing the arrow keys
Default all alarms are set to NO, which means they are not activated; activating the alarms or not is
the choice of the user; alarms do not influence the output and purity.
2. When you select YES or AUTO RESET, the rest of the alarm parameters that need to be set will pop-up
automatically (see screen below).