K3.1.142k Manual NitroFlow Lab - 24 -
Replace O2 cell When an O
2
cell has been replaced during maintenance, this button can be
touched and a new date to replace the O
2
cell is set. System asks for
confirmation. In maintenance screen 1/5 the O
2
lifetime should read 3 year ahead
from date of changing.
6.3.7 Data logging menu
Access: Press data logging menu button in settings screen
(refer to § 6.3)
Function: Read the logged (saved) data
Alarms as well as status of the sensors are saved on the SD-card. The time between the logging (saving) of this
data to the memory card can be chosen in the data-logging menu.
Button Selection Result
Interval 30-3600 Time in seconds between the logging (saving) of alarm
data
Shows all the alarms that have been saved on the memory
drive (see below)
Button Explanation
A Date and time of alarm incident
B Alarm description
C The number of logged alarms
CAUTION:
Please check the alarm, p-switch, options and settings before you restart the unit.
The unit cannot run without the SD-card. This will generate an alarm (SD-card failed).