K3.1.142k Manual NitroFlow Lab - 16 -
6 Operation of the control system
6.1 Menu structure
The menu structure of the control system is built up as shown below. One can always go back to a higher level in
the menu by pushing the
-button.
6.2 Main screen
Access: This is the start-up screen that automatically appears when the generator is switched on.
Function: Gives access to the different menus.
Symbol/data Information/result
When flashing there is an alarm; touch the symbol and the current alarm will
be shown.
Status of unit (A) Can be OFF/RUN/STAND-BY/ALARM/P-RELIEF
Menu button, touch to go to settings menu
Switch ON/OFF button, generator will turn ON or OFF
To turn the unit on, touch the switch -button. The status will switch to P-RELIEF. The compressors will start three
minutes (180 seconds) after the unit has been switched on. The delay time countdown time is shown next to the
text P-RELIEF (see below).
A
Main screen
Alarm
display
Settings Switch on/off unit
Data
logging
Log on Set
alarms
Pressure
switch
Options Local
settings
Maintenance