K3.1.142k Manual NitroFlow Lab - 22 -
Button Selection Result
Language English, Francais, Deutsch,
Nederlands, Español
Text in the screen will appear in the chosen
language.
Pressure BAR/PSI* Pressure indications will appear in the
chosen setting
Flow LPM/CFM Flow will appear in the chosen setting
Purity %N2/%O2 Purity will appear in nitrogen (%N2) or
oxygen (%O2) percentage
* Select BAR, to display temperature in °C. Select PSI to display temperature in °F.
6.3.6 Maintenance menu
Access: Touch maintenance menu button in settings screen
(refer to § 6.3)
Function: Shows maintenance status and offers calibration possibility.
The maintenance menu consists of 5 different screens. Each screen displays maintenance status or calibration
buttons.
SCREEN 1/5
Data Explanation
Type Shows type of generator this unit is
Version Software revision number
O2 lifetime Month-year when O
2
-sensor needs to be changed (3 years from data of order)
Filter life Hours countdown from 1 year to 0 hrs
SCREEN 2/5
Data Explanation
Total Total running hours of the generator