K3.1.142k Manual NitroFlow Lab - 19 -
Button Selection Result
Active No Alarm function for this parameter is not active
Active Yes Alarm function for this parameter is active; alarm messages must
be reset manually
Active Auto reset Alarm function for this parameter is active; When alarm level is
not exceeded any longer before manual reset, the alarm will reset
itself
Stop Yes Generator will switch off in case alarm level is exceeded
Stop No An alarm signal will be given but generator will continue to run in
case alarm level is exceeded
Level 0-16% O
2
100 – 84% N
2
For screen 1/6 and 2/6: this is the oxygen- or nitrogen level* at
which the alarm is set.
Level 0-13 BAR*
0-188.5 PSI*
For screen 3/6 and 4/6.
This is the pressure level at which the alarm is set
Level 0-10 BAR*
0-145 PSI*
For screen 5/6 and 6/6.
This is the pressure level at which the alarm will appear.
Delay 0-300 sec Delay time in seconds between the moment that the alarm level
has been exceeded and signaling; this feature prevents false
alarms in case of short spikes
*see also local settings -menu
ATTENTION:
It is impossible to set O
2
low at a higher level than O
2
high. The setting of O
2
low is limited once
O
2
high has been set. Therefore, first set O
2
high level before setting the O
2
low level.