7
2. SPECIFICATIONS
Item Specification FOR USA & CANADA EUROPE & OCEANIA
Power Requirements Voltage 120 VAC (NTSC MODEL) 230V (PAL MODEL)
Frequency 50 60 Hz
Power consumption 1.5A 0.8A
Voltage fluctuation +/– 10 %
Operating Environment Temperature 10 ~ 40°C
Relative humidity 30 ~ 85 %
Air pressure 700-1060 hPa
Storage/transport Temperature –20 ~ 60°C
Environment Relative humidity 0 ~ 85 %
Air pressure 700 ~ 1060 hPa
Illumination Lamp Metal Halide Lamp
MSCR 70-50E
Lamps average life span 200 hours, continuous use
Color temperature 12,000 K
Lighting format Switching regulator with continuous illumination
Brightness control Selection - Automatic or Manual
Automatic iris Servo type
Scope Compatibility Pentax color video endoscopes
NTSC MODEL PAL MODEL
Pentax fiberscopes With use of appropriate fiberscope video adapter module
NTSC MODEL PAL MODEL
Other manufacturers’ fiberscopes With use of appropriate fiberscope video adapter module and
appropriate eyepiece/light guide adapters
Air Feed System
Air pump system Electro-magnetic vibrator system
Pressure setting at flow rate of 0 45 ~ 49 kPa 41 ~ 62 kPa
Standard air feed volume at inlet of water bottle
3.1 ~ 7.0L/min 3.2 ~ 8.0L/min
Water Feed System Water bottle assembly pressurized by pump
Bottle capacity = 250mL
Brightness Control System
Automatic Selection - Average or Center
Manual +/– 5 step adjustment
Color System Color correction Red +/– 5 steps
Blue +/– 5 steps
Freeze Function Live video image provided
when freeze mode activated
Cooling Forced air cooling
Video Outputs 2 sets: RGBS (NTSC or PAL), 9pin D-Sub female connectors
2 sets: Composite (NTSC or PAL), BNC connector
2 sets: Separate Video (Y/C), 4-pin female connector
1 set : Computer, 9-pin D-Sub female connector
Classification as Type of protection against electric shock Class I equipment, 3 prong plug
Electro Medical Equipment
Degree of protection against electric shock BF Type (Body Floating), using insulated endoscope.
Use on heart is prohibited
Degree of explosion proofing Do not use in potentially flammable surroundings
Acoustic Noise Sound pressure level < 70 dB (A) (Based on ISO 7779)
Electro Magnetic Compatihility
CISPR11 IEC 60601-1-2
Compliance Designed in accordance with
UL 2601-1
IEC 60601-1
Size Dimensions Width = 380 mm Height = 153 mm Depth = 405 mm
Weight Main Body = 13.5 kg (29.8 lbs)