Pentax EPK-700 - Water Bottle Assembly

Pentax EPK-700
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11
3-2 WATER BOTTLE ASSEMBLY
No. NAME FUNCTION
1 Bottle Holds sterile water for procedure. (use up to 2/3’s full)
2 Water Bottle Cap Assembly Must be firmly secured to bottle to prevent air leakage.
3 Air Pipe Stem Inserts into video processor water bottle receptacle.
4Air/Water Hose Contains (2) independent tubes - 1 for air, 1 for water.
5 Air/Water Connector Inserts into Air/Water socket of endoscope umbilical connector.
6Air/Water- Drain Lever Must be set to upright (A/W) position for delivery of air and water.
7 Water Feeding Stem Channel for water to be displaced from bottle and into scope.
1
2
3
6
7
4
5
- If the water bottle assembly has been handled roughly, the water feeding tube inside the Air/Water hose may be
disconnected at the A/W connector for the endoscope. To test, remove the cap assembly and using a syringe, inject water into the
water feeding stem. If the water comes out of both the center hole of the A/W connector and the series of holes around the center
hole, the water feeding tube is disconnected. Use another water bottle assembly.
NOTE