11
3-2 WATER BOTTLE ASSEMBLY
No. NAME FUNCTION
1 Bottle Holds sterile water for procedure. (use up to 2/3’s full)
2 Water Bottle Cap Assembly Must be firmly secured to bottle to prevent air leakage.
3 Air Pipe Stem Inserts into video processor water bottle receptacle.
4Air/Water Hose Contains (2) independent tubes - 1 for air, 1 for water.
5 Air/Water Connector Inserts into Air/Water socket of endoscope umbilical connector.
6Air/Water- Drain Lever Must be set to upright (A/W) position for delivery of air and water.
7 Water Feeding Stem Channel for water to be displaced from bottle and into scope.
- If the water bottle assembly has been handled roughly, the water feeding tube inside the Air/Water hose may be
disconnected at the A/W connector for the endoscope. To test, remove the cap assembly and using a syringe, inject water into the
water feeding stem. If the water comes out of both the center hole of the A/W connector and the series of holes around the center
hole, the water feeding tube is disconnected. Use another water bottle assembly.