6.3 Error when no atmospheric correction is made
When measurement is carried out with no Atmospheric Correction (with the settings fixed at a
temperature of 15°C and an atmospheric pressure of 1013 hPa or 760 mmHg), the Error per 100
meters in temperature and pressure will be shown in the tables below.
When the actual pressure is 1013 hPa (760 mmHg) and the temperature is 25°C, conducting
the measurement with the temperature left at 15°C will result in the measurement being
short by 0.9 mm per 100 meters.
[Error table: When hPa (15°C, 1013hPa as standard)]
[Error table: With mmHg (15°C, 760mmHg as standard)]