26
14. Occlude green port and observe that pulsations stop.
15. Remove occlusion from green port.
16. Press the MANUAL INSP button and observe pressure increase over
40 cmH₂O.
NOTE: If using the “EUR” multimeter, MAP is in a dierent position on display.
17. Rotate the MASTER switch to the “OFF” position.
18. Disconnect gas source supply. Pre-use check is complete.