33
4. Twist white cap on the rear of the
Phasitron®5 to remove.
5. Remove cap.
6. Remove sliding venturi with spring
from Phasitron®5 body.
Cleaning Phasitron®5
1. Thoroughly rinse each of the disassembled parts (except for tubing
harness and lter) under warm running tap water for approximately
10 seconds.
2. Use fragrance-free liquid soap, added to a clean bowl or basin with
warm water.
3. Hand wash all parts of the Phasitron®5 kit and accessories in the
warm soapy water.
4. Rinse all parts thoroughly using sterile water.
5. Gently shake all parts to remove as much water as possible, then dry
with a clean lint-free cloth or paper towel.
6. Use a clean, damp cloth to wipe the exterior of the tubing harness
with an approved alcohol based cleaner.
7. Reassemble the Phasitron®5 and place in the supplied bag until next use.
8. Do not disinfect the Phasitron®5 for reuse with more than one patient.