EasyManua.ls Logo

PERDIX SHEARWATER - Getting Started; Basic Setup; 4.1. Install the Transmitter

PERDIX SHEARWATER
20 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Instructions
Page 6
RevA
FIGURE 5 Install transmitter on 1st stage HP port
Install transmitter on the same side of your body as the
handset. Range is approximately 3 feet (1 m).
FIGURE 4 A backup SPG is recommended
4. Getting Started / Basic Setup
This section will get you started with the basics of AI on the
Perdix� Advanced setup and detailed descriptions will be
covered in later sections�
4.1. Install the Transmitter
Before using the AI system, you will need to install one or
more transmitters on a scuba tank rst stage regulator.
The transmitter must be installed on a rst stage port labeled
“HP” (high pressure). Use a rst stage regulator with at least
two HP ports, so that a backup analog submersible pressure
gauge (SPG (FIGURE 4)) can be used�
Position the transmitter such that it is on the same side of your
body as you wear your Perdix AI handset (FIGURE 5)� Range is
limited to approximately 3 ft (1 m)�
A high-pressure hose may be used to relocate the transmitter
for better reception or convenience� Use hoses rated for a
working pressure of 4500 PSI (300 Bar) or higher
Range
3 ft (1 m)
Range
3 ft (1 m)
Use a wrench (5/8”) to tighten or
loosen the transmitter
Avoid hand tightening or loosening,
as this can stress the body of the
transmitter