EasyManua.ls Logo

PERDIX SHEARWATER - 4.2. Turn On the Transmitter; 4.3. Turn Off the Transmitter; 4.4. Enable Al on the Perdix

PERDIX SHEARWATER
20 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Instructions
Page 7
RevA
4.2. Turn On the Transmitter
Turn on the transmitter by opening the tank valve� The
transmitter will automatically wake up when it detects
pressure�
Pressure data is transmitted every 5 seconds�
4.3. Turn Off the Transmitter
To turn o󰀨 the transmitter, close the tank valve and purge the
second stage regulator to drain pressure from the hoses� The
transmitter will automatically power down after 30 seconds of
no applied pressure�
Leave the valve open and the transmitter on for now
4.4. Enable AI on the Perdix
On the Perdix AI, navigate to the System SetupAI Setup
menu (FIGURE 6)� Change the AI Mode setting to T1 (Tank 1)�
The AI is now on�
When AI Mode is Off, the AI sub-system is completely
powered down and does not consume any power� When on
the AI system increases power consumption by approximately
10%� For example, an alkaline AA battery which lasts
approximately 45 hours with AI o󰀨 (on medium screen
brightness), would last about 40 hours with AI on�
More information about settings in the AI Setup menu can be
found in Section 5.1. AI Setup
AI Mode
Off
Change Save
AI Setup
AI Mode
GTR Mode
Units
T1 Setup
T1
T1
PSI
000000
Change Save
AI Setup
FIGURE 6 Enable AI by changing the AI Mode to T1
The menu above can be found at System SetupAI Setup