Clarus SQ 8 MS Hardware Guide
114
Cleaning Materials
Wooden stick cotton swabs
Deionized Water
6000 Grade Micro Mesh (Part No. N9303420)
8000 Grade Micro Mesh (Part No. N9303421)
600 grit aluminum oxide in DI Water with a few drops of methanol to make a
paste
Acetone
Methanol
When cleaning internal components it is important to maintain the quality of the
surface finish. Deep scratches or pits can cause loss of performance. Where no specific
cleaning procedure is provided, you should use fine abrasives to remove dirt from
metal components. Recommended abrasives are:
6000 Grade Micro Mesh (Part No. N9303420).
8000 Grade Micro Mesh (Part No. N9303421).
600 grit aluminum oxide in DI Water with a few drops of methanol to make a paste
After cleaning with abrasives, it is necessary to wash all metal components in suitable
solvents to remove all traces of grease, oil and, if micro-mesh is used, rubber. The
recommended procedure is to swill or sonicate the components in a clean beaker of
methanol for at least ten minutes and subsequently to blot them dry with lint-free
tissue. Recommended solvents are:
After the components are reassembled, they should be blown with oil-free nitrogen to
remove dust particles.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PerkinElmer CLARUS SQ 8 MS and is the answer not in the manual?

PerkinElmer CLARUS SQ 8 MS Specifications

General IconGeneral
BrandPerkinElmer
ModelCLARUS SQ 8 MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals