Warnings and Safety Information
Power Requirements
Power Consumption Clarus MS: 1000 VA (volt-amps)
Clarus GC: 2400 VA (volt-amps)
Add 100 VA for the computer and 108 VA for a printer.
Power Consumption
(with optional oven
heater)
See below listing for the Clarus 580/680. Add 100 VA for the
computer and 108 VA for the printer.
Power Specification All electrical supplies must be smooth, clean, and free of line
transients greater than 40 V peak to peak, and must meet and
remain within the following tolerances:
Clarus MS:
120 VAC ±10 % @ 50/60 Hz ±1 % 1000 VA maximum
230 VAC ±10 % @ 50/60 Hz ±1 % 1000 VA maximum
Clarus GC:
For GC with slow heating rate as standard;
120 VAC ± 10% @ 50/60 Hz ± 1% @ 20 Amps, 2400 VA
maximum
230 VAC ± 10% @ 50/60 Hz ± 1% @ 10 Amps, 2400 VA
maximum
For GC with optional oven heater for fast heating rate;
220 VAC ± 5% @ 50/60 Hz ± 1% @ 15 Amps, 3120 VA
maximum
230 VAC ± 5% @ 50/60 Hz ± 1% @ 16 Amps, 3120 VA
maximum
240 VAC ± 5% @ 50/60 Hz ± 1% @ 13 or 16 Amps, 3120
VA max
Instruments and peripherals should not be connected to
circuits with large inductive or large and frequent loads (for
example, large motors, discharge lamps, photocopy systems,
radio transmitters, etc.).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PerkinElmer CLARUS SQ 8 MS and is the answer not in the manual?

PerkinElmer CLARUS SQ 8 MS Specifications

General IconGeneral
BrandPerkinElmer
ModelCLARUS SQ 8 MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals