42 Lambda 265, 365 and 465 ASX-260/280 AND ASX-520/560 Autosampler Users Guide
Wavelength Monitoring Measurement Procedure
NOTE:
In case of Lambda 465, start the sample measurement after warming up the system at least
20 minutes.
1. Launch the UV Lab Software and then select Wavelength Monitoring and click OK.
2. After setting parameters for Experiment Setup, Baseline Correction and Wavelength
Monitoring, click OK.
3. Select Auto Sampler in the Accessory Type and click setup button (Pencil
icon) in the Experiment Setup on the main screen of the software.
4. Select the autosampler model. Click OK.