UV Lab Software. 51
8. Click Blank.
9. The probe will move to the blank position and start sucking and, after filling up the
flow cell, the blank will be measured.
10. Click Standard.
NOTE:
User can set up to 10 standard samples in the Autosampler. The number of standards
should be matched with the number of concentration entry in the Quantification Standard
set-up windows.
11. The probe will move to the first set standard position and start sucking and after
filling up the flow cell and the standard will be measured. The standard
measurement will continue until the last standard measurement is completed.
12. After all standards measurements, click Quantification Sample mode.
13. Click Blank.
14. The probe will move to the blank position and start sucking and, after filling up the
flow cell, the blank will be measured.
15. Click Sample.
16. The probe will move to the first set sample position and start sucking and after filling
up the flow cell and the sample will be measured. The sample measurement will
continue until the last sample measurement is completed.
17. Save or print the results after finished.