The effects of increasing differences with
increasing moisture contents can be
corrected by means of the multiplicative
factor GAIN. For this purpose at least two
samples, if possible, of different moisture
contents are required, since otherwise
the effects cannot be determined
exactly.
Since the procedure is complicated
(determination of the gain of a straight
line) we recommend to consult the
manufacturer before changing the GAIN!
Change the factor by means of the arrow
keys and terminate the input by pressing:
Select again »Offset« and the next
product to be changed, etc.
»»<----« select with the arrow keys
and press in order to return to the
main menu.
Now select the menu item »Off«:
and switch off by pressing .
Settings
Off
OFF
Barley
Offset: -0.5
Gain : 1.000