Page 35 E
VERFLO
TM
SERVICE & TECHNICAL INFORMATION 1039055, VER. 06
SYMPTOM CAUSE VERIFICATION
CORRECTIVE
ACTION
LED failure
No power to LED’s
Verify the power cord
is plugged in
Verify power in wall
outlet
Verify that all wiring
harnesses are
connected
Verify the Main PCA
is not damaged
Connect the
power cord to AC
power source
Check the
household fuse or
circuit breaker
Connect all wiring
harnesses
Replace the Main
PCA
Unit is inoperable
Main PCA failure
Verify the wiring
harnesses are
connected
Verify the wiring
harness is not
damaged
Verify the condition
of the fuse on PCA
Verify the condition
of the Power Cord
Connect the wiring
harnesses
Check the wiring
harnesses for
continuity
Test the fuse for
electrical
continuity
Verify the Power
Cord continuity
Replace Power
Cord
Replace the Main
PCA
Start-up alarms in-
operable
Piezoelectric speaker
failure
Verify the solder
joints on the
piezoelectric speaker
Replace the Main
PCA