SERVICE & TECHNICAL INFORMATION 1039055, VER. 06
SYMPTOM CAUSE VERIFICATION
CORRECTIVE
ACTION
Low oxygen output
Sieve bed failure
Compressor failure
Verify the oxygen
output with a
calibrated oxygen
analyzer
Verify the
Compressor is
working properly
Verify that there are
no leaks
Verify Compressor is
operating properly
Replace the Sieve
Canister
Replace the
Compressor
Replace tubing
and clamps
Replace
Compressor
Low Oxygen
percentage
indicator not
functioning
Continuous red
LED
Main PCA failure
Verify the OPI output
Replace the Main
PCA
Warm cabinet
temperature
Cooling Fan failure
Verify the Cooling
Fan connectors are
seated properly
Verify there is
continuity on
electrical fan wiring
harness
Verify there is power
to the Cooling Fan
Verify the Cooling
Fan is not obstructed
Reseat the
connectors
Replace the fan
assembly
Reseat the
connectors
Remove the
obstructions from
the Cooling Fan