Page 37 E
VERFLO
TM
SERVICE & TECHNICAL INFORMATION 1039055, VER. 06
SYMPTOM CAUSE VERIFICATION
CORRECTIVE
ACTION
Pressure relief
valve activated
Red LED (constant)
Low pressure
Cycle failure
Verify Valve Solenoid
Assembly operation
Verify Valve Solenoid
assembly wire
harness connection
with Main PCA
Verify voltage from
Main PCA to Valve
Solenoid Assembly
Replace Valve
Solenoid Assembly
Reconnect the Valve
Solenoid wiring
harness to the Main
PCA
Replace the Main PCA
Inaccurate oxygen
readings
Leaks
Main PCA failure
Compressor failure
Verify the hoses are
connected to the OPI
sensor.
Verify that there are
no leaks
Verify the output with
a calibrated oxygen
analyzer
Verify the
Compressor output
range.
Reconnect the
hoses
Reconnect the
tubing
Replace Main PCA
Replace the
Compressor