Page 64 E
VERFLO
TM
SERVICE & TECHNICAL INFORMATION 1039055, VER. 06
3.
Secure the Valve Solenoid Assembly to the Sieve Canister Assembly by securing the five screws
to 8 in-lbs. using the torque sequence listed in Refer to Figure 8-19.
4.
Replace cable ties, ensuring the cable ties are facing downward towards the valve body, as
shown below.
CAUTION
The washer is important so the screw in this location does not penetrate a channel in the valve.