Page 72 E
VERFLO
TM
SERVICE & TECHNICAL INFORMATION 1039055, VER. 06
Figure 8-26: COMPRESSOR PLACEMENT INTO FRONT CABINET
2.
Connect the blue hose to the Sieve Canister Assembly and secure it using a new one-eared
clamp.
3.
Connect the black Compressor wires to the Capacitor.
4.
Place the Compressor springs in the grooves at the top of the Compressor Mounting area.
5.
Lock the Compressor in place by sliding the mounting rod though the compressor springs and into
place.
6.
Install the Main PCA. Refer to the Main PCA Replacement section for more information.
7.
Install the Rear Cabinet. Refer to the Rear Cabinet/Power Cord Replacement section for more
information.
8.
Install the Inlet Filter. Refer to the Inlet Filter Replacement section for more information.
9.
Install the Filter Cover. Refer to the Filter Cover Replacement section for more information.
NOTE
The Philips Respironics manufactured compressors are not repairable and must be replaced.
NOTE
Devices manufactured after 06/18/2013 can be repaired with GSE Compressors. See Section 9.2 for RP
Kit P/Ns. See Section 8.9 for a Serial No. cutoff for various EverFlo models. Devices originally equipped
with short (11.5 inch) sieve canisters may not have the GSE compressor installed as a replacement.
To replace compressor tubing and foam
1. Remove the faulty blue tubing from the compressor outlet by cutting the clamp.