Performance Assurance Tests - Checking and Calibrating the Gas Analyzer 4 Testing and Maintenance
55
NOTE
If your PC/Laptop does not have an RS232 serial connector, you can use a USB-to-RS232 Serial
adapter. For the COM port settings, you need to change the USB Serial port to COM1. To do this:
1 Go to Control Panel -> System -> Hardware - Device Manager and select
the USB Serial Port.
2 Right click on the USB Serial Port, click Properties and select Advanced.
3 Select COM1 from the drop down menu and click OK.