Recommendations for Test Frequency
Testin
g
the Instrument
139
Recommendations for Test Frequency
The testing checklist appears in the next section of this chapter. The listed procedures should
be performed as indicated in the Suggested Testing Timetable below. The checklist may be
photocopied and should be completed by the tester. It should be filed for future reference.
Suggested Testing Timetable Frequency
Functionality Assurance
Performance Assurance Test
System Check
System Self-Test
To verify proper operation when
the monitor has been installed.
an assembly has been replaced.
to verify the operation any time the
unit is suspect.
Preventive Maintenance Tests
NBP Calibration
Sidestream CO
2
Calibration
Once every two.
Once a year, or following any
Instrument repairs or the replacement
of any Instrument parts (applies to
units with serial number prefix
DE020xxxxx only).
Performance and Safety Tests
Temperature Accuracy
ECG/Resp Performance
Invasive Pressure Performance
• SpO
2
Performance
• Mainstream CO
2
Performance
Nurse Call Relay Performance
ECG Sync Performance
Once every two years.
including Safety Checks (in
accordance with IEC 60601-1)
System Enclosure Leakage
Current
Protective Earth
Patient Leakage Current
Once every two years.