Functionality Assurance Tests
Testin
the Instrument
145
System Self-Test Values
Module Test Numeric Limits Test Waveform
ECG and ECG/
Resp
ECG
Resp
100 bpm in ADULT mode
125 bpm in NEO/PEDI mode
15 rpm in ADULT mode
30 rpm in PEDI mode
55 rpm in NEO mode
Simulated squarewave and
numeric
Simulated Resp wave and numeric
Pressure
a
Systolic
Diastolic
a. An appropriate transducer must be plugged into the Server or Server Extension in
order to get the test reading.
120 mmHg ADULT
60 mmHg PEDI/NEO
0 mmHg ADULT/PEDI/
NEO
Simulated squarewave and
numerics
SpO
2
/Pleth
a
SpO
2
Pleth
777
9
100%
Wave on display
Numeric only
Simulated wave
Temperature
a
40 °C (104 °F) Numeric only
Pleth Pulse Rate 60 bpm in ADULT/NEO
mode
NBP 120/80 (90) ADULT
100/60 (80) PEDI
80/50 (60) NEO
Simulated squarewave and
numerics
Sidestream CO
2
b
b. Appropriate tubing must be plugged into the Server Extension in order to get the
test reading. When in Neonatal mode the test signals are switched on, a “CO
2
Equip. Malf.” message will appear until the test signals are switched off; no
action is required.
EtCO
2
40 mmHg
ImCO
2
0 mmHg
AWRR 20 rpm
Simulated squarewave and
numerics
Mainstream CO
2
a
EtCO
2
40 mmHg
ImCO
2
0 mmHg
AWRR 20 rpm
Simulated squarewave and
numerics