Power supply
ac
mains
reference value
220 V
nominal values
110 V/128 V/220 V/238 V,
selectable
by solder links
nominal
operating range
±10
% of
selected nominal value
operating limits
±10
%
of selected nominal value
nominal frequency
range
50-100
Hz
limit range of
operation
47,5
-
105 Hz
power consumption
13 W
Environmental capabilities
The following
environmental
data are valid
only if the
instrument
is
checked
in
accordance with
the
official checking
procedure.
Details on these
procedures
and failure criteria are
supplied on request by
the PHILIPS
organization in
your country or
by Philips International B.V., Industrial & Electro-acoustic
Systems
Division, Eindhoven,
The Netherlands.
Ambient
temperature:
reference value
+23
°C
±1 K
nominal working
range
+
5
°C
...
+40
°C
limit range
of
operation
+
5 °C ...
+40
°C
limits for storage and transport
-40
°C
...
+70
°C
Relative humidity:
reference range
45 ... 75 %
nominal working
range 20
... 80 %
limit
range of operation
10... 85%
limits
for
storage
and transport
0...
85%
Air pressure:-
reference value
1000
±15 hPa
nominal
working range
798... 1064 hPa
Airspeed:'
reference value-
0 ... 0.2 m/s
nominal working range
0 ... 0.5 m/s
Heat
radiation: direct sunlight radiation
not allowed
Vibration:
limits for storage and
transport max. 0.35 mm amplitude
(10 to 60
Hz)
max.
5
g
(60 to 150 Hz)
radio interference
voltage level
of
interference
< K
operating position
normally upright
on feet or with handle
fold down
warm-up time
5
min
Cabinet
protection type {see DIN 40
050) IP 20
protection
class
{see
IEC
348)
class
1,
protective conductor
line
connection
overall
dimensions:
mains cable,
fixed
to the
instrument
height
140 mm
width
310
mm
depth
310 mm
weight
4.8
kg (11
lb)