E
1-6
1.3.
1.3.1.
1.3.2.
•>
ACCESSORIES
Standard accessories
operating manual
fuse and
voltage labels
2-terminal
test fixture
Optional accessories
service
manual
4
-wire
test cable
RCL adapter
with
2
single
test posts
and 1 double test post
SMD test fixture
9499
520 08201
5322 265
24026,
Fig.
33
By
means of the
2
-terminal test
fixture
common compo-
nents are connected
9499 525 00911
PM 9541
,
Fig. 34
PM
9542, Fig. 35
PM 9542
SMD, Fig. 36
optional accessory for
PM 9542
For precise results
low-ohmic impedances < 100 £2 should be
measured applying
4
-wire system.
For
this the
4-wire test cable with Kelvin clamps
PM
9541
and the RCL
adapter PM
9542
are available.
PM 9541 is also ideal for
in-circuit testing of components.
PM 9542 is
designed
to
provide
rapid low impedance connection to
the instrument,
whatever
the shape
and dimension
of the component under test.
Please
remove the
two single test posts, if
you
use
the
double
test post for
CUT
connection
or remove
the double
test
post
if
you use the
two
single test
posts.
Wrong insertion
of the plug
into the RCL meter is prevented
by unsymmetrical
arrangement
of the
pins.
About compensation
of
the zero capacitance chapter 3.4.3. gives some
information.
When measuring
low-ohmic
components with PM
9541,
the short-circuit
inductance
of
max.
0.3
juH
of
the
cable must be
taken into account.
PM 9542
SMD is designed
for measurement of
SMD components (Surface
Mounted
Devices).
The SMD
test
fixture is
used together
with the RCL adapter PM 9542.
For understanding
the measurement
circuit when applying the
accessories, see figs.
32
-
36.
Performance
characteristics
p"*-
,
A
i
PM 9541
PM 9542
PM
9542
SMD
j
1)
CUT connection
2 Kelvin clips
Kelvin contacts
within the
test posts
Kelvin
contacts
^
J
•
short-circuit inductance
0,1
MH,
max.
0.3
/tH
<
0.1 /2H
<
0.1
/iH
. I i
:
1
i
:
\
1
f
1
: .(
measuring
accuracy
with PM 6303
dimensions of
components under
test
-
length
—
width
—
height
mechanical
specifications
as for PM
6303,
but
additional
error for
very
low-ohmic CUTs
caused by the internal
short-circuit inductanco
as for PM 6303
as for PM
6303
but additional error for
CUTs with very
low
capacitance,
see chap. 3.4.5. and
Fig.
36a
min.
2 mm/max.
10
mm
min. 1
mm
min.
0.5
mm
i
—
cable length
0.6
m 0.6
m
f
—
dimensions
50
mm x
1
45
mm
x 95 mm 45
mm
x 55 mm
x 30 mm
i
J
—
weight
0.2 kg
0.6 kg
0.03
kg
|
1
•
1
A
;
spare
parts:
none
double test post:
none
5322 264 30185;
2 single
test posts (red
and black);
5322 264 30184