47
SimplyGo User Manual
English
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
This device is intended for use in the electromagnetic environment specied belowThe user of this device should assure that it
is used in such an environment�
Immunity Test IEC 60601 Test Level
Compliance
Level Electromagnetic Environment - Guidance
Portable and mobile RF communications equipment should be
used no closer to any part of the Device, including cables, than the
recommended separation distance
Conducted RF
IEC 61000-4-6
3 V
150 kHz to 80 MHz
6 V in ISM & amateur
radio bands between
150kHz and 80 MHz
3 V
6 V
Recommended separation distance
Minimum 30 cm (12 inches)
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2�7 GHz
Up to 28 V/m in
telecommunications
bands as specied
in clause 8�10 of
IEC60601-1-2:2014
10 V/m
28 V/m
Interference may occur in the vicinity of equipment marked
with the following symbol: