Philips RESPIRONICS System One - Home Cleaning Instructions; Humidifier Enclosure, Heater Plate and Exterior Outlet Port; Outlet Port Interior; Water Tank and Lid

Philips RESPIRONICS System One
31 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13
Home cleaning Home Disinfection Hospital cleaning Hospital Disinfection
Humidier
Door Seal
Immerse and agitate
in a solution of mild
liquid dishwashing
detergent (e.g. Dawn
Ultra) and potable
water.
Frequency: Daily for
1 year (366 cycles
maximum)
Immerse and agitate
in 70% Isopropyl
Alcohol.
Frequency: Every 3
days for 1 year (122
cycles maximum)
Immerse and agitate
in a commercially
available enzymatic
detergent solution
(e.g. Medizime LF).
Frequency: Prior to
high-level disinfec-
tion
Immerse in a 75±2°C bath of sterile
deionized water for 30 minutes.
Frequency: Every 3 days and be-
tween patients for 1 year (122 cycles
maximum)
Dry Box, Dry
Box Seal
and Inlet
Seal
Immerse and agitate
in a solution of mild
liquid dishwashing
detergent (e.g. Dawn
Ultra) and potable
water.
Frequency: Daily for
1 year (366 cycles
maximum)
Immerse and agitate
in 70% Isopropyl
Alcohol.
Frequency: Every 3
days for 6 months
(60 cycles maximum)
Immerse and agitate
in a commercially
available enzymatic
detergent solution
(e.g. Medizime LF).
Frequency: Prior to
high-level disinfec-
tion
Immerse in a 75±2°C bath of sterile
deionized water for 30 minutes.
Frequency: Every 3 days and between
patients for 6 months (60 cycles
maximum)
Heated
Tubing
Immerse and agitate
in a solution of mild
liquid dishwashing
detergent (e.g. Dawn
Ultra) and potable
water.
Frequency: Every 3
days for 6 months (60
cycles maximum)
Immerse and agitate
in 70% Isopropyl
Alcohol.
Frequency: Every 3
days for 6 months
(60 cycles maximum)
Immerse and agitate
in a commercially
available enzymatic
detergent solution
(e.g. Medizime LF).
Frequency: Prior to
high-level disinfec-
tion
Immerse in a 75±2°C bath of sterile
deionized water for 30 minutes.
Frequency: Every 3 days and between
patients for 6 months (60 cycles
maximum)
Immerse and agitate in Cidex OPA
Frequency: Every 3 days and between
patients for 6 months (60 cycles
maximum)
Home Cleaning Instructions
Humidifier Enclosure, Heater Plate and Exterior Outlet Port
WARNING: Allow the humidier heater plate and water to cool down for approximately 15
minutes before removing the water tank. A burn may result from touching the heater plate,
coming in contact with the heated water, or touching the tank pan.
WARNING: Before cleaning the humidier, always remove the humidier from the therapy
device.
1. Open the humidier lid with the release lever, and then slide the water tank out of the humidier
enclosure.
2. Clean the humidier enclosure, heater plate and outlet port exterior by wiping it for one min-
ute with a lint-free cloth dampened with a solution of mild liquid dish washing detergent (e.g.
Dawn Ultra) and potable water. A ratio of 1 teaspoon of mild liquid dish washing detergent per
gallon (3.8 liters) of water is required.
3. Wipe with a new lint-free cloth dampened (not dripping) with potable water for one minute,
turning the cloth frequently, to remove all detergent residue. Allow to air dry completely be-
fore reconnecting to the therapy device.
4. Inspect the humidier enclosure for any damage and replace it if necessary.