Instruction and Operator's Manual PICO pH
11
Some key points:
Perform calibration for accuracy greater than 0.2 pH
This operation should be carried out at the beginning of each season when the pool is put into
service, and then preferably every 3 to 4 months to ensure that the pH reading remains as
accurate as possible.
The procedure should be repeated in the event of abnormal chemical consumption .
3.2 Pump priming
3.3 Settings
3.3.1 Information
PICO pH function
Factory setting
Unit of measure
pH
Setpoint definition
From pH 7.0 to 7.6
Regulation type
Acid - reduce pH by dosing an acid (pH-)
Calibration to pH 7
To be carried out for accuracy greater than 0.2 pH
Proportional regulation *
Factory-set function
* Proportional: Dosing modulated through distinct on/off cycles. When approaching the setpoint, the pump reduces its
dosing time.
The start-up of PICO pH- is now complete, and it is ready to measure the pH of your pool.
1) Dive the electrode 3
minutes in the pH7
solution, stir, then leave
it in the solution without
touching either the
electrode or the cable
(single-use solution).
2) Once the value has stabilized, press CAL until “CAL” flashes,
then release. Next, an indication of the probe quality between
0 and 100% is briefly displayed.
With the unit off (display OFF), press and hold the
ON/OFF button for at least 3 seconds to start priming.