USER MANUAL Page 56 of 69
When the notation is released, the red light turns off.
The Medical Diode Laser monitors the operating status of the system in real time.
Through Sounds, indicators, and graphical warning signs alert user happened
unusual events and interrupting the laser outputting.
1.Laser temperature exceeds
45 ° C;
2. Fiber coupler temperature
exceeds 60 ℃;
3. Temperature sensor failure;
1. Turn off the power, if the temperature
notation caused by the high temperature,
air cooling device for 10 minutes and it will
return to normal after the boot;
2. Please contact the company after-sales
maintenance.
The remote interlock is not
inserted correctly or is not
plugged into the device;
1.Check the rear panel of the device and
whether the remote interlock is correctly
inserted into the device;
2. If the remote interlock is inserted into the
device and the notation condition is still
unresolved, please contact the company
for after-sales maintenance.
1.The foot switch is not properly
inserted or is not inserted into
the device;
2.Press the handle switch button
when it’s not into the "Ready"
state;
1.Check the front panel and whether the
handpiece switch is correctly inserted into
the device;
2.If the handle switch is properly inserted
into the device, the notation status is still
unresolved, please check the handle switch
button is pressed or not;
3. Check the above operation, if there is still
notation status, please contact the
company after-sales maintenance.
Please contact the company for after-sales
maintenance.
1.Fiber connector is not inserted
into the laser window;
2.Micro - switch of laser window
fails;
1.Insert the fiber connector into the laser
window of the device;
2.If there is still a notation condition, please
contact the company after-sales
maintenance.
Use external power supply